ANALYSIS OF ACCURACY AND MEASUREMENTS OF FLOWING MEDIUM PARAMETERS BY MEANS OF A WAVE THERMOANEMOMETER SYSTEM

Zygmunt Biernacki, Tadeusz Zloto, Marek Kurkowski, Pawel Ptak
Abstract:
The paper analyses the measurement accuracy of selected parameters, such as flow velocity and temperature, of a flowing medium, by means of the thermoanemometric method. The core of the method relies on the Wave Thermoanemometer System constructed by one of the authors. The sources of boundary error components are analysed within the signal conversion channels of the WTS. Errors are examined by means of relevant mathematical expressions representing signal conversion both in the velocity measuring channel and in temperature measuring channel. The algorithms have been developed for perfect and for real conditions. The theoretical considerations have been verified by laboratory tests performed by computerized equipment.
Keywords:
flow velocity, wave thermoanemometer, measurement accuracy, signal conversion
Download:
IMEKO-TC4-2002-039.pdf
DOI:
-
Event details
IMEKO TC:
TC4
Event name:
TC4 Conference and Workshop 2002
Title:
4th International Conference on Advanced A/D and D/A Conversion Techniques and their Applications (together with 7th IMEKO TC4 Workshop on ADC Modelling and Testing)
Place:
Prague, CZECH REPUBLIC
Time:
26 June 2002 - 28 June 2002