Use of a 100 MΩ step Hamon resistor for calibration of pico-ammeters in dc current in the range 100 pA ÷ 100 nA |
---|
Pier Paolo Capra, Flavio Galliana, Milena Astrua, E. Gasparotto |
- Abstract:
- This paper describes a measurement circuit for the calibration of pico-ammeters in dc current. The system is based on a 100 MΩ Hamon network developed at INRIM to improve the traceability level of the maintained 1 GΩ standard. Besides the Hamon network, the measurement system, presented in this paper, consists of a precision dc voltage source and an electronic circuit used as voltage guard driver.
- Download:
- IMEKO-TC4-2007-012.pdf
- DOI:
- -
- Event details
- IMEKO TC:
- TC4
- Event name:
- TC4 Symposium 2007
- Title:
- XVth IMEKO TC4 International Symposium on Novelties in Electrical Measurements and Instrumentation (together with 12th IMEKO TC4 Workshop on ADC Modelling and Testing)
- Place:
- Iasi, ROMANIA
- Time:
- 19 September 2007 - 21 September 2007