Temperature influence on the frequency response of the Keysight 3458A digital multimeter
Y. A. Sanmamed, J. R. Salinas, J. Díaz de Aguilar, F. García-Lagos, R. Caballero
-
Abstract:This paper describes the procedure and experimental set-up used at CEM to determine the temperature influence on the frequency response of two Keysight 3458A digital multimeters (DMMs) in the Direct Current Volt (DCV) sampling mode. The DMMs were used to digitalize an AC input signal of 0.8 V and 1 kHz provided by an AC source.
-
Keywords:Analog to Digital Converter, temperature coefficient, aperture time, digital sampling, digitizer
-
Download:
-
DOI:_unreg_tc4-2019.042
Event details:
[EVENTDETAILS]
-
IMEKO TC:TC4
-
Event name:TC4 Symposium 2019
-
Title:
23rd IMEKO TC4 Symposium
-
Place:Xi'an, CHINA
-
Time:17 September 2019 - 20 September 2019