Skip to main content

REALISATION OF A PRIMARY AIR VELOCITY STANDRAD USING LASER DOPPLER ANEMOMETER AND PRECISION WIND TUNNEL

Jian Wu
  • Abstract:
    A primary air velocity meter calibration standard comprising of a precisely controlled wind tunnel system and a Laser Doppler Anemometry (LDA) system was developed for the calibration of air velocity meters from 0.2 m/s to 60 m/s with measurement uncertainties of less than 0.18% to 0.63%. The LDA system is used to verify the wind speed in the test section of the wind tunnel, and provide traceability to the SI for air speed measurements.
  • Keywords:
    air velocity measurement, wind profiles, measurement uncertainty
  • DOI:
    _unreg_wc-2009.413

Event details:

[EVENTDETAILS]
  • IMEKO TC:
  • Event name:
    XIX IMEKO World Congress
  • Title:

    Fundamental and Applied Metrology

  • Place:
    Lisbon, PORTUGAL
  • Time:
    6 September 2009 - 11 September 2009