Skip to main content

Negative Step Dynamic Torque Calibration Machine

Tao Li, Xianglong Ma, Zhiqiang Qiao
  • Abstract:
    The paper develops a negative step dynamic torque calibration machine and illustrates its working principle and critical technologies. The use of retracting actuators reduces significantly the fall time of the negative step torque, whereas air bearing yields high transfer accuracy of the dynamic torque. The experimental study is conducted and the results show that the fall time of the step torque reaches to millisecond level and the dynamic response up to 5800 Hz can be excited. Through uncertainty analysis, the expanded uncertainty of the presented calibration machine is better than 5 % (k = 2).
  • Keywords:
    negative step torque, calibration machine, retracting actuator,air bearing
  • DOI:
    _unreg_tc3-2017.034

Event details:

[EVENTDETAILS]
  • IMEKO TC:
    TC3
  • Event name:
    TC3 Conference 2017 - Measurement facing new challenges!
  • Title:

    23rd Conference on the Measurement of Force, Mass and Torque (together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness)

  • Place:
    Helsinki, FINLAND
  • Time:
    30 May 2017 - 1 June 2017