Abstract:
The paper describes the design, measurement results and uncertainty analyses of the hydraulic driven piston-prover system which is in operation at VSL since 2008. The 12 meter long, 0.6 m bore piston-prover is used for the realization of Reference Values for Gas-Volume at pressures between 1 and 65 bar(a) and ‘any’ type of Gas. The principle is based on the displacement of a piston acting as a Gas-Oil separator. The standard has a flow-rate range from 5 to 230 m³/h. The system is designed to calibrate reference meters. The Calibration and Measurement Capability (CMC) of the system is proven smaller than 0.1% (k = 2). The paper also explains the coherence between the Gas-Oil Piston-Prover and other traceability generators and ‘flow rate bootstrapper systems’.
-
Keywords:gas, oil, piston-prover, traceability
-