Search Proceedings
Who we are
CALIBRATION OF MULTI-DECADE INDUCTIVE VOLTAGE DIVIDERS
J. Bohacek, M. Andrle
-
Abstract:A method for calibrating multi-decade inductive voltage dividers has been developed which makes it possible to determine both the inphase and the quadrature error for ratios of i /11, i = 1, 2, ..., 10 in the frequency range from 400 Hz to 5 kHz. Calibration is made by comparing the divider under test against a reference two-stage divider having a ratio winding with 11 identical sections. In calibration of the reference divider itself, an auxiliary 11:1 transformer is used.
-
Keywords:inductive voltage divider, calibration, ratio error
-
Download:
-
DOI:_unreg_wc-2000.216
Event details:
[EVENTDETAILS]
-
IMEKO TC:
-
Event name:XVI IMEKO World Congress
-
Title:
Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future
-
Place:Vienna, AUSTRIA
-
Time:25 September 2000 - 28 September 2000