A LOW COST IMPLEMENTATION OF THE SETTLING-TUBE MEASUREMENT TECHNIQUE FOR SAND GRANULOMETRY

A. Papalazarou, K. Albanakis, Th. Laopoulos
  • Abstract:
    The settling-tube measurement technique is implemented by an experimental setup developed and tested for low-cost automated sand granulometry. The measurement is based on monitoring very small pressure variations created by samples of sand as they settle down in a water-filled cylinder. The pressure variations are monitored by low cost air-pressure differential transducers in an appropriate experimental setup which monitors air pressure variations caused by water pressure variations. It is shown that this low-cost automated approach manages to calculate properly size distribution and other statistical parameters of sand samples efficiently and quickly offering thus an attractive alternative to the traditional sheaving method.
  • DOI:
    _unreg_tc19-2010.112

Event details:

[EVENTDETAILS]
  • IMEKO TC:
    TC19
  • Event name:
    Instrumentation for the Information and Communication Technology Era
  • Title:
    3rd IMEKO TC19 Symposium (together with 17th IMEKO TC4 Symposium & 15th International Workshop on ADC Modelling and Testing)
  • Place:
    Kosice, SLOVAKIA
  • Time:
    8 September 2010 - 10 September 2010