Abstract:Accurate pH measurement is vital in wastewater treatment for effective process control and regulatory compliance. This study compares a self-cleaning pH sensor (HORIBA 6122 series) with a conventional glass electrode, installed in parallel at a wastewater treatment facility in Metro Manila, Philippines. Over a 97-day period, 2,324 paired readings were recorded. Statistical analysis revealed a significant difference (t(2323) = 56.88, p > 0.05) in mean values (HORIBA = 6.93, conventional = 6.12). The HORIBA sensor exhibited lower variability (HORIBA = 0.2335, conventional = 0.6461) through time, indicating better stability. These results align with the goals of Process Analytical Technology and support the deployment of self-cleaning sensors for real-time wastewater treatment monitoring.
-