Search Proceedings
Who we are
Influence of Coordinate System Alignment on the Calibration of Multi-Component Force and Moment Sensors
J. Nitsche, D. Röske, R. Tutsch
-
Abstract:For the calibration of multi-component force and moment sensors no national or international standard exists. In different approaches for calibration procedures, a precise alignment of sensor and calibration reference is required. The influence of such an alignment, however, is not mentioned.
Based on a model of a calibration setup and an assumption of possible misalignments, an analysis of the systematic error, caused by such misalignment, is performed. Additionally, a strategy to identify the necessary coordinate transformation is given. -
Keywords:force, moment, multi-component, calibration
-
Download:
-
DOI:_unreg_tc3-2017.017
Event details:
[EVENTDETAILS]
-
IMEKO TC:TC3
-
Event name:TC3 Conference 2017 - Measurement facing new challenges!
-
Title:
23rd Conference on the Measurement of Force, Mass and Torque (together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness)
-
Place:Helsinki, FINLAND
-
Time:30 May 2017 - 1 June 2017