EMISSION RATE CORRECTIONS FOR 241AmBe SOURCES IN MANGANESE BATH OF LNMRI/IRD
Sandro P. Leite, Walsan W. Pereira, Evaldo S. da Fonseca, Karla C. de Souza Patrão
-
Abstract:This work tries to get better: uncertainties and accuracy for determination of bath losses. With simulation is possible to individualize calculations from all the loss components, for each specific measurement. A regression model is also evaluated based on an old and a new proposed model.
-
Keywords:manganese bath, monte carlo simulation, neutron metrology, neutron sources
-