Search Proceedings
Who we are
Air Density Determination By Using Buoyancy Artefacts at SASO NMCC
M. S. Alghamdi, R. A. Alyousefi, S. M. Alqarni, A. A. Aljuwayr, S. Kacmaz
-
Abstract:In this study, the air density has been determined with relative uncertainty of 1,75 × 10-5 (k = 1) by using experimental method to establish SASO NMCC’s mass scale at the highest accuracy. Hollow and tabular type buoyancy artefacts, which have the same surface area, but a different volume of 279, 1196 cm3, have been compared under the vacuum of 5 × 10-6 mbar and in air by using 1 kg vacuum mass comparator.
-
Keywords:air density, buoyancy artefacts, vacuum mass comparator
-
Download:
-
DOI:_unreg_tc3-2017.061
Event details:
[EVENTDETAILS]
-
IMEKO TC:TC3
-
Event name:TC3 Conference 2017 - Measurement facing new challenges!
-
Title:
23rd Conference on the Measurement of Force, Mass and Torque (together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness)
-
Place:Helsinki, FINLAND
-
Time:30 May 2017 - 1 June 2017